| Cable | Adapter | Port | VR Device Name |
|---|---|---|---|
| Direct serial cable | None | COM1 | Medibus |
If COM1 is already in use, see When the COM1 Port is Already in Use below.
For the connection shown below, use a Y-cable with a receive-only branch to Vital Recorder. Enable Read Only Mode when adding the device.
| Machine end — DB-9M | Existing connection — DB-9F (CON1) | PC end — DB-9F (CON2) |
|---|---|---|
| 2 (TxD from machine) | 2 | 2 (RxD at PC) |
| 3 (RxD to machine) | 3 | Not connected |
| 5 (GND) | 5 | 5 (GND) |
Vital Recorder records the data requested by the existing device. If waveform order is not detected correctly, set wavs= to match that device's requested waveform order.
No changes are required if COM1 is already configured as follows. Otherwise, apply these settings in the machine's serial interface menu; the menu path varies by model and software version.
| Parameter | Value |
|---|---|
| Protocol | MEDIBUS |
| Baud Rate | 9600 |
| Data Bits | 8 |
| Parity | Even |
| Stop Bits | 1 |
Add the device as Medibus and select the PC serial port used for the connection.
For a Y-cable connection, enable Read Only Mode. For a direct connection, wavs= can be used to request specific waveforms, such as wavs=AWP,AWF for airway pressure and flow.