Note: This Infinity Acute Care System (IACS) connection records numeric data through the P2500 export port. ECG and arterial pressure waveforms require the Analog/Sync connection and an ADC.
| Connection | Cable or Interface | Adapter | Port | VR Device Name |
|---|---|---|---|---|
| Numeric data | Dräger export protocol cable MS22948 or custom RJ-10 to DB-9F cable | None for the cable described below | Export port on P2500 | Infinity |
| ECG/BP waveforms | Compatible Analog/Sync cable, connection cable for the ADC, and ADC | Depends on the ADC connection | Analog/Sync connection on M540 | Select the device type for your ADC |
Use the P2500 export port for this serial connection. No additional null modem adapter is required with the custom pin assignment below.
For analog waveforms, use a compatible Dräger Analog/Sync cable and ADC. If an Infinity M-Cable Microstream CO2 is also connected to the M540, use the compatible Y-cable arrangement for those accessories.
| RJ-10 Pin (P2500 end) | DB-9F Pin (PC end) |
|---|---|
| 3 | 2 |
| 2 | 3 |
| 4 | 5 |
Waveforms are only available as analog voltages, read through an ADC (SNU-ADC, DataQ DI-149/DI-155, …).
No changes to the monitor settings are required for the numeric-data connection described here.
Infinity and select the PC serial port used for the connection.