Note: The 170 Series and 250cx use different serial connectors and setup procedures. Use the section for your model.
| Model | Cable | Adapter | Port | VR Device Name |
|---|---|---|---|---|
| 170 Series | Custom RJ-45 to DB-9F cable with RTS–CTS loop | None | RS-232 Port 1 or Port 2 | Coro |
| 250cx | Compatible RJ-11 serial cable | Confirm for the cable used | RS-232C: J109, J110, or J111 | Coro |
The custom cross cable provides the TX/RX crossover, so no separate Null Modem adapter is used.
Monitor RJ-45 port:
| RJ-45 pin | Port 1 signal | Port 2 signal | Direction |
|---|---|---|---|
| 1 | +5 V (200 mA, fused) | GND | — |
| 2 | RTS | RTS | Output from monitor |
| 3 | RXD | RXD | Input to monitor |
| 4 | GND | GND | — |
| 5 | GND | GND | — |
| 6 | TXD | TXD | Output from monitor |
| 7 | CTS | CTS | Input to monitor |
| 8 | +5 V (200 mA, fused) | GND | — |
Pins 1 and 8 of Port 1 carry +5 V and are not connected to the PC side.
Cable wiring — three data conductors, crossed:
| RJ-45 (monitor) | Signal | DB-9F (PC side) | Signal |
|---|---|---|---|
| 6 | TXD | 2 | RX |
| 3 | RXD | 3 | TX |
| 4 or 5 | GND | 5 | GND |
Connect RJ-45 pin 7 (CTS) to RJ-45 pin 2 (RTS) at the monitor end.
The 250cx has RJ-11 RS-232C ports (J109, J110 and J111), not the 170 Series' RJ-45 ports, so it needs a separate compatible RJ-11 serial cable. Confirm whether an adapter is required for the cable used. The 170 Series pinout above does not apply.
The communication mode and baud rate for each port live in service setup mode, which can only be entered from a power-off state.
± qualifier is lit; the FHR display is active when the heartbeat indicator is lit. | Setup code (UA display) | Parameter | Value (FHR display) | |---|---|--- | | 30 | RS-232 Port 1 — communications mode | 5 (= 115 update) | | 31 | RS-232 Port 1 — baud rate | 9600 | | 40 | RS-232 Port 2 — communications mode | 5 (= 115 update) | | 41 | RS-232 Port 2 — baud rate | 9600 |
0 = HP, 1 = HP w/notes, 2 = ext. BP, 3 = factory test, 4 = ext. FSpO2, 5 = 115 update, 6 = 115 transmit/receive.96.Coro.30/40 select the communications mode for ports 1/2 and 31/41 set their baud rate.