Note: The Atlan speaks Dräger MEDIBUS.X and is added in Vital Recorder as
MedibusX. The COM port must be set to MEDIBUS.X at 19200 before data arrives.
| Cable | Adapter | Port | VR Device Name |
|---|---|---|---|
| Direct serial cable | Null modem adapter (F/F) | COM1 or COM2 | MedibusX |
Open the machine's COM interface settings and configure the connected port as follows. Refer to the model's instructions for the menu path.
| Parameter | Value |
|---|---|
| Protocol | MEDIBUS.X |
| Baud Rate | 19200 |
| Data Bits | 8 |
| Parity | Even |
| Stop Bits | 1 |
Add the device as MedibusX and select the PC serial port used for the connection.