Drager Perseus

Drager Perseus

Note: Password-protected interface configuration required. See Device Configuration below.

CableAdapterPortVR Device Name
Direct SerialNull Modem F/FCOM1 or COM2MedibusX

Connection Steps

  1. Attach a Null Modem (F/F) adapter to the serial port.
  2. Connect a direct serial cable from the adapter to the PC via USB-Serial converter.

Device Configuration

  1. Navigate to System Settings → System Menu. Enter password 0000.
  2. Select Interface Configuration.
  3. For the connected port (COM1 or COM2), set Protocol → MEDIBUS and Baud Rate → 9600.