Note: This guide describes the connection for Aisys CS2 and Avance CS2. Confirm the connector and pinout with GE for other models.
| Cable | Adapter | Port | VR Device Name |
|---|---|---|---|
| Custom 9-pin ↔ 15-pin serial | None | 15-pin female connector | Datex-Ohmeda |
| Machine side — DB-15 male | PC side — DB-9 female |
|---|---|
| 13 — TX | 2 — RX |
| 6 — RX | 3 — TX |
| 5 — GND | 5 — GND |
For the connection shown below, use a Y-cable with a receive-only branch to Vital Recorder. Enable Read Only Mode when adding the device.
| Machine end — 15-pin D-sub male | Existing connection — 15-pin D-sub female (CON1) | PC end — DB-9F (CON2) |
|---|---|---|
| 13 (TxD from machine) | 13 | 2 (RxD at PC) |
| 6 (RxD to machine) | 6 | Not connected |
| 5 (GND) | 5 | 5 (GND) |
Only the machine's transmit signal and ground connect to the PC branch.
No changes to the machine settings are required for this setup.
Add the device as Datex-Ohmeda and select the PC serial port used for the connection.
For a Y-cable connection, enable Read Only Mode. Available data depends on the existing communication between the machine and the connected device.