Hamilton G5 Ventilator
Note: Available in Vital Recorder v1.10.2 or later. Either Monitoring Interface port (1 or 2) can be used.
| Cable | Adapter | Port | Protocol | VR Device Name |
|---|
| Direct Serial | Null Modem M/F | Monitoring Interface 1 or 2 | Block Protocol | Hamilton |
Connection Steps
- Attach a Null Modem (M/F) to either Monitoring Interface port on the rear.
- Connect a direct serial cable from the adapter to the PC via USB-Serial converter.
Device Configuration
Configuration is only accessible when the device is not currently operating.
- Press the two designated buttons simultaneously to open the Configuration menu (bottom left of screen).
- Press the two Test Mode activation buttons simultaneously to enter Test mode.
- Navigate to Configuration → Interface.
- Set the connected port (COM1 or COM2) to "Block Protocol".
- Press Close → Close/Save.