Medtronic (Aspect Medical) BIS A-2000

Medtronic (Aspect Medical) BIS A-2000

Note: Medtronic BIS A2000 is similar to BIS Vista, but BIS A2000 can obtain 2-channel 256Hz EEG.

CableAdapterPortVR Device Name
direct serial cableNoneRS-232 portA2000

Connection Steps

  1. Connect a direct serial cable to the RS-232 serial port on the lower right of the rear panel.
BIS A-2000 rear panel — DB-9 J1 RS-232 serial port circled, below the model label (MONITOR Model A-2000, P/N 185-0070) and beside the AC power inlet

Device Configuration

The serial protocol lives several levels deep, in the diagnostic/service area of the menu tree. Navigate with the front-panel keys.

  1. Press Menu to open the Setup Menu, then select Advanced Setup (bottom row).
A-2000 Setup Menu — Event, Sensor Check, Display Type, BIS Smoothing Rate, with Advanced Setup highlighted on the bottom row
  1. In the Advanced Setup Menu, select Diagnostic Menu.
A-2000 Advanced Setup Menu — Secondary Variable, Time/Date, Print Event, Display Parameter Setup, with Diagnostic Menu highlighted; Save Settings and Return to Setup Menu at the bottom
  1. In the Diagnostic Menu, select System Configuration Menu.
A-2000 Diagnostic Menu — DSC Self Test, Display Self Test, Sensor Data Display, Clear Data, Diagnostic Codes, Impedance Checking, with System Configuration Menu highlighted
  1. On Serial Port Protocol choose Binary.
  2. Press Return To Diagnostic Menu → Return to Advanced Setup Menu → Save Settings. Without Save Settings, the protocol reverts on the next power cycle.
A-2000 System Configuration Menu — Serial Port Protocol row with ASCII / Binary options, Binary selected; Extended Memory and ICU Mode rows below