Edwards Lifesciences EV-1000 / EV-1000A

Edwards Lifesciences EV-1000 / EV-1000A

Note: The adapter gender differs between the two generations. The old EV-1000 (label REF EV1000M) has a male DB-9, so it needs a Null Modem adapter (F/F). The newer EV-1000A (label REF EV1000A) has a female DB-9 and needs a Null Modem adapter (M/F). Check the REF label on the rear panel before ordering the adapter.

ModelCableAdapterPortVR Device Name
EV-1000 (REF EV1000M)Direct serial cableNull modem adapter (F/F)Rear DB-9 male serial portEV1000
EV-1000A (REF EV1000A)Direct serial cableNull modem adapter (M/F)Rear DB-9 female serial portEV1000

Connection Steps

  1. Check the rear REF label and serial connector. Connect a direct serial cable using the null modem adapter listed for that model.

EV-1000 (REF EV1000M):

 <img src="../hardware_images/edwards_ev1000_1.png" width="450" alt="Old EV-1000 (REF EV1000M) rear panel with the male DB-9 serial port circled, sited between the USB ports and the VGA connector">
  • EV-1000A: the serial connector is a female DB-9 immediately to the right of the port marked LAN2. Attach a Null Modem adapter (M/F).
 <img src="../hardware_images/edwards_ev1000_2.png" width="450" alt="EV-1000A rear panel with the female DB-9 serial port circled, immediately right of the LAN2 port">

Device Configuration

  1. Select the Settings icon.
EV-1000 monitoring screen with the gear-shaped settings icon in the left icon strip circled
  1. Select Monitor Settings.
EV-1000 Settings menu showing Patient Data, Monitor Settings, Parameter Settings, Data Download, Demo Mode, Engineering and Help, with Monitor Settings circled
  1. Select Serial Port Setup.
EV-1000 Monitor Settings menu showing General, Date/Time, Monitoring Screens, Serial Port Setup, Restore All Defaults and Advanced Features, with Serial Port Setup circled
  1. Set Device to IFMout.
EV-1000 Serial Port Setup screen with Device set to IFMout circled, alongside Stop Bits 1, Data Bits 8, Parity None and Flow Control 2 seconds
  1. Set Baud Rate to 9600 and confirm the remaining settings below.
EV-1000 baud rate picker listing 9600, 19200, 38400 and 57600, with 9600 circled
ParameterValue
DeviceIFMout
Baud Rate9600
ParityNone
Stop Bits1
Data Bits8
Flow Control2 seconds

Vital Recorder Setup

Add the device as EV1000 and select the PC serial port used for the connection.