LiDCO (LiDCOplus / LiDCOrapid)
Note: Serial output is off by default. The RS-232 communications module must be switched on inside the monitor's Engineering Screen before Vital Recorder will see anything, and the baud rate here is 57600 — not the 9600 used by the Edwards monitors.
| Cable | Adapter | Port | Serial | VR Device Name |
|---|
| direct serial cable | Null Modem adapter (M/F) (assumes female COM1; unverified) | COM1 — standard 9-way serial port on the monitor | 57600 baud | LiDCO |
Connection Steps
- Attach a Null Modem adapter (M/F) to COM1, the standard 9-way serial port on the monitor.
- Connect a direct serial cable from the adapter to the PC via a USB-Serial converter. The Null Modem supplies the TX/RX crossover; a direct serial cable alone will not work.
Device Configuration
- Enable the serial link and set its parameters in the monitor's Engineering Screen. The communications software module must also be switched on. The menu path is Settings → Communications → Serial.
- Set the following:
| Parameter | Value | |-----------|------- | | LiDCO Serial | Enabled | | Baud Rate | 57600 | | Data Bits | 8 | | Parity | None | | Stop Bits | 1 | | Average | Never | | Observation | Beat-to-beat |
- Average = Never and Observation = Beat-to-beat matter as much as the baud rate: with averaging enabled the monitor emits smoothed values at a slow cadence instead of per-beat data, and Vital Recorder's trend will look sparse or frozen.
Vital Recorder Setup
- In the Vital Recorder device list this appears under Cardiac monitor as LiDCO: LiDCO.
Troubleshooting
Settings → Communications → Serial is not in the menu. Menu wording varies between LiDCOplus and LiDCOrapid and across software revisions. Look for the RS-232 port configuration inside the Engineering Screen.
- Port opens but no data arrives. Check the adapter against the COM1 connector. If the port is male, use a Null Modem adapter (F/F). Confirm that the serial values match the table above.