Note: The EVITA family speaks Dräger MEDIBUS.X. The volume waveform is not transmitted, and CO2 requires a capnography module.
| Cable | Adapter | Port | VR Device Name |
|---|---|---|---|
| direct serial cable | Null modem; connector gender not confirmed | RS-232 COM1 (or COM2) | MedibusX |
Open the COM port / interface settings and configure the connected port as follows. The menu path varies by model and software version.
| Parameter | Value |
|---|---|
| Protocol | MEDIBUS.X |
| Baud Rate | 19200 |
| Data Bits | 8 |
| Parity | Even |
| Stop Bits | 1 |
MedibusX and select the PC serial port used for the connection.