Hamilton G5 / C-series Ventilators (G5, C6, C2, T1, MR1)
Note: Either Monitoring Interface port (1 or 2) can be used, but the port's protocol must be switched to Block first.
| Cable | Adapter | Port | VR Device Name |
|---|
| direct serial cable | Null Modem (M/F) | Monitoring Interface 1 or 2 | Hamilton |
Connection Steps
- Find the connector column on the side/rear of the ventilator. Three connectors are stacked there:
- Monitoring Interface 1 connector (9-pin) = COM1
- Monitoring Interface 2 connector (9-pin) = COM2
- Attach a Null Modem adapter (M/F) to Monitoring Interface 1 or 2.
- Connect a direct serial cable from the adapter to the PC via a USB-Serial converter.
The COM (RS-232) connector pin assignment is: 2 RxD, 3 TxD, 4 DTR, 5 GND, 6 DSR, 7 RTS, 8 CTS (pins 1 and 9 unused, shield = chassis ground).
Device Configuration
The interface controls only appear when the ventilator is in Standby (not ventilating a patient).
- Enter Configuration mode: press the Oâ‚‚ enrichment and Manual breath keys at the same time. A Configuration button appears at the bottom left of the screen.
- Enable Test mode: press the Screen Lock/Unlock and Nebulizer On/Off keys at the same time. Both Configuration and Test modes must be enabled for the interface controls to be editable — but do not actually enter Test mode.
- Touch Configuration, then Interface.
- On the block for the port you cabled (COM1 protocol or COM2 protocol), select
HAMILTON-G5 / Block.
- Touch Close, then Close/Save to store the setting.
- Block protocol line settings: 38400 baud, 8 data bits, No parity, 1 stop bit, no handshake
- Do not select
HAMILTON-G5 / Polling or Galileo / Polling — polling runs at 9600 baud, 7 data bits, Even parity, 2 stop bits with XON/XANY handshake and transmits only a subset of the data. DraegerTestProtocol is for Dräger MIB II converters and HAMILTON-G5 / Block (ACK) for distributed alarm systems.
Vital Recorder Setup
- In Vital Recorder, add the device as
Hamilton.