Maquet / Getinge Servo-i / Servo-s / Servo-U Ventilators

Maquet / Getinge Servo-i / Servo-s / Servo-U Ventilators

Note: On Servo-i and Servo-s, connect to the BOTTOM RS-232 port. The TOP port is for service/debugging. The live port and adapter requirement for Servo-U have not been verified; confirm both with Getinge before connecting.

CableAdapterPortSerialVR Device Name
direct serial cableServo-i / Servo-s: Null Modem adapter (M/F); Servo-U: unverifiedServo-i / Servo-s: BOTTOM RS-232; Servo-U: unverifiedServo-i / Servo-s: 9600 baud, Even parity (7 or 8 data bits, 1 or 2 stop bits โ€” auto-detected), XON/XOFF ยท Servo-U: 19200Servo-i

Connection Steps

  1. Open the connector cover on the rear of the patient unit. On Servo-i and Servo-s, identify the BOTTOM of the two ports labeled RS 232; the upper port is for service/debugging.
  2. On Servo-i and Servo-s, attach a Null Modem adapter (M/F) to the bottom port.
  3. Connect a direct serial cable from the adapter to the PC via a USB-Serial converter.

Device Configuration

Nothing has to be enabled on the ventilator. The serial port is served by the built-in Computer Interface Emulator (CIE), which answers commands from Vital Recorder as soon as the line is up.

Model differences: Servo-i and Servo-s run the CIE at 9600 baud; Servo-U runs at 19200 baud. Set the matching device entry in Vital Recorder.

Vital Recorder Setup