Masimo Radical-7

Masimo Radical-7

CableAdapterPortVR Device Name
direct serial cableNoneP1 (RS-232)Radical7

Connection Steps

  1. Connect a direct serial cable to P1 (RS-232) on the rear of the Docking Station.
Docking Station rear panel — P1 marked RS-232 next to the wider P2 nurse-call/analog connector
  1. Connect a direct serial cable from P1 to the PC's serial port. A USB-Serial converter can also be plugged straight into P1 — no Null Modem adapter is required.
  2. Confirm that the handheld is fully docked before recording.

Device Configuration

  1. From the main menu, select DEVICE SETTINGS.
Radical-7 main menu — SOUNDS, DEVICE SETTINGS, ABOUT, 3D
  1. The device settings screen offers ACCESS CONTROL and DEVICE OUTPUT. Select DEVICE OUTPUT.
Radical-7 device settings screen — ACCESS CONTROL and DEVICE OUTPUT tiles
  1. Set serial → ASCII 1, and analog 1 and analog 2 → Pleth.
Device output screen — serial set to ASCII 1, analog 1 and analog 2 set to Pleth
  1. Scroll down and set docking station baud rate → 9600. SatShare diagnostics may be left Enable.
Device output screen scrolled down — SatShare diagnostics Enable and docking station baud rate 9600