Sentec SDM

Sentec SDM

Note: The serial interface is disabled while trend data are being downloaded over the LAN port. It is reactivated automatically once the LAN download finishes.

CableAdapterPortVR Device Name
Direct serial cableNoneRS-232 portSDM

Cable Pinout

DB-9 PinSignal
2TxD (from SDM)
3RxD (to SDM)
5Signal ground
1, 4, 6–9Reserved

Connection Steps

  1. Connect a direct serial cable to the Serial Data Port (RS-232) on the rear of the SDM.
Sentec SDM rear panel — the RS-232 serial data port among the surrounding interface connectors
  1. Connect the other end to the PC through a USB-Serial converter.
  2. If using a USB-Serial converter, connect its USB end to the PC.

Device Configuration

Open Interfaces → Serial Interface and apply the settings below for this Vital Recorder connection.

ParameterValue
ProtocolSenTecLink
Baud Rate115200
Serial Interface menu with protocol set to SenTecLink and baud rate 115200

Vital Recorder Setup