Nihon Kohden BSM

Nihon Kohden BSM

Note: The availability of RS-232C and ECG/BP OUT ports depends on the monitor model and installed interface or input unit. If the required port is not present, contact Nihon Kohden to confirm the compatible hardware and installation options. RS-232C supports numeric data only. ECG and invasive blood pressure waveforms require an ECG/BP OUT connection and an analog-to-digital converter (ADC).

ConnectionCable or InterfaceAdapterDevice PortVital Recorder Device Type
Numeric data — QI-373PDirect serial cable (DB-9 M/F)Null modem (M/F)RS-232C on the QI-373P interfaceBSM
Numeric data — other interfacesCable appropriate for the interfaceConfirm for the interfaceRS-232CBSM
ECG/BP waveformsCompatible Nihon Kohden ECG/BP output cable, connection cable for the ADC, and ADCDepends on the ADC connectionECG/BP OUTSelect the device type for your ADC

Connection Requirements

Numeric Data

ModelRS-232C Interface
BSM-1700 seriesBuilt-in; no additional interface board required
BSM-3000 seriesQI-373P, subject to model compatibility
BSM-6301QI-631P
BSM-6501 / BSM-6701QI-671P

The direct serial cable + null modem adapter (M/F) configuration applies to QI-373P. Confirm the cable and adapter requirements for other interfaces.

ECG/BP Waveforms

ModelECG/BP OUT
BSM-1700 seriesBuilt into the monitor; no additional interface board required
BSM-3000 seriesProvided by a compatible QI-371P or QI-372P interface
BSM-6000 seriesProvided by a compatible AY input unit; not available on AY-660P

An RS-232C interface does not necessarily provide ECG/BP OUT.

Use a compatible YJ-910P or YJ-920P ECG/BP output cable and an ADC. The connection between the output cable and the ADC depends on the cable termination and the ADC's input connector.

Connection Steps

Numeric Data

  1. Identify the RS-232C socket (DB-9 female, marked with the serial IOIOI icon) on the interface unit's panel. Do not confuse it with the 15-pin RGB/video socket (IOI) directly below it, or with the RJ-45 network socket.
BSM interface unit connector panel with the DB-9 RS-232C socket outlined in red, above the 15-pin video socket, the RJ-45 network socket and the ECG/BP OUT connector
  1. Attach a Null Modem adapter (M/F) to that socket. Secure the adapter with the connector screws.
  2. Connect a direct serial cable from the adapter to the PC's DB-9M port or a USB-Serial converter.

ECG and Arterial Pressure Waveforms

ECG and arterial pressure waveforms come out of the ECG/BP OUT port as analog voltages, visible on the same connector panel as the serial socket.

  1. Plug the Nihon Kohden ECG/BP output cable into the ECG/BP OUT port.
  2. Build a 5.5pi Mono ↔ RJ45 cable to bring the analog outputs into the ADC (SNU-ADC / SNUADCM, DataQ DI-149/DI-155, …).
  3. Connect the ADC to the PC via USB.

Network Collection Through a Central Station

Where a Nihon Kohden central station with the HL7 gateway is installed, Vital Recorder can collect data over the network without a per-bed serial cable. Add these three devices:

Vital Recorder deviceDataPort
NIHONKOHDEN::ADTPatient ID9007
NIHONKOHDEN::ORFNumeric data every 30 seconds7999
NIHONKOHDEN::NealTimeWaveforms, up to three — default ECG_II, PLETH and AWP9001

Network collection may be available through a compatible Nihon Kohden central monitoring system and HL7 gateway. Confirm the available data and connection settings with Nihon Kohden and the hospital's system administrator.

Device Configuration

No changes to the monitor settings are required.

Vital Recorder Setup