Note: The availability of RS-232C and ECG/BP OUT ports depends on the monitor model and installed interface or input unit. If the required port is not present, contact Nihon Kohden to confirm the compatible hardware and installation options. RS-232C supports numeric data only. ECG and invasive blood pressure waveforms require an ECG/BP OUT connection and an analog-to-digital converter (ADC).
| Connection | Cable or Interface | Adapter | Device Port | Vital Recorder Device Type |
|---|---|---|---|---|
| Numeric data — QI-373P | Direct serial cable (DB-9 M/F) | Null modem (M/F) | RS-232C on the QI-373P interface | BSM |
| Numeric data — other interfaces | Cable appropriate for the interface | Confirm for the interface | RS-232C | BSM |
| ECG/BP waveforms | Compatible Nihon Kohden ECG/BP output cable, connection cable for the ADC, and ADC | Depends on the ADC connection | ECG/BP OUT | Select the device type for your ADC |
| Model | RS-232C Interface |
|---|---|
| BSM-1700 series | Built-in; no additional interface board required |
| BSM-3000 series | QI-373P, subject to model compatibility |
| BSM-6301 | QI-631P |
| BSM-6501 / BSM-6701 | QI-671P |
The direct serial cable + null modem adapter (M/F) configuration applies to QI-373P. Confirm the cable and adapter requirements for other interfaces.
| Model | ECG/BP OUT |
|---|---|
| BSM-1700 series | Built into the monitor; no additional interface board required |
| BSM-3000 series | Provided by a compatible QI-371P or QI-372P interface |
| BSM-6000 series | Provided by a compatible AY input unit; not available on AY-660P |
An RS-232C interface does not necessarily provide ECG/BP OUT.
Use a compatible YJ-910P or YJ-920P ECG/BP output cable and an ADC. The connection between the output cable and the ADC depends on the cable termination and the ADC's input connector.
IOIOI icon) on the interface unit's panel. Do not confuse it with the 15-pin RGB/video socket (IOI) directly below it, or with the RJ-45 network socket.
ECG and arterial pressure waveforms come out of the ECG/BP OUT port as analog voltages, visible on the same connector panel as the serial socket.
ECG/BP OUT port.Where a Nihon Kohden central station with the HL7 gateway is installed, Vital Recorder can collect data over the network without a per-bed serial cable. Add these three devices:
| Vital Recorder device | Data | Port |
|---|---|---|
NIHONKOHDEN::ADT | Patient ID | 9007 |
NIHONKOHDEN::ORF | Numeric data every 30 seconds | 7999 |
NIHONKOHDEN::NealTime | Waveforms, up to three — default ECG_II, PLETH and AWP | 9001 |
Network collection may be available through a compatible Nihon Kohden central monitoring system and HL7 gateway. Confirm the available data and connection settings with Nihon Kohden and the hospital's system administrator.
No changes to the monitor settings are required.
BSM and select the PC serial port used for the connection.