Fresenius Kabi Link+ Agilia

Fresenius Kabi Link+ Agilia

Note: Data is recorded through USB. Enable Data Export through the LAN connection before recording.

CableAdapterPortVR Device Name
USB 2.0 A-male ↔ mini-B 5-pinNoneUSB-B (mini-USB)Link+

Connection Steps

  1. Connect the mini-B end to that port and the USB-A end directly to the PC. No USB-Serial converter is needed — the rack presents itself as a USB CDC-ACM serial device.

Device Configuration

Link+ connector panel with the mini-USB (USB-B) data port highlighted, above the USB-A socket and beside the RJ-45 and round connectors

Enabling Data Export is a one-time procedure performed over LAN from a PC. It must be done before the USB connection will produce any data.

  1. Connect the Link+ LAN (RJ-45) port — on the left of the same connector panel — to the PC's Ethernet port with the LAN cable supplied with the Link+.
Link+ connector panel with the RJ-45 LAN port highlighted on the left side of the panel
  1. On the PC, open Control Panel → Network and Internet → Network Connections, right-click the Ethernet adapter and choose Properties.
Windows Network Connections with the Ethernet adapter right-clicked and Properties highlighted in the context menu
  1. In the adapter properties list, select Internet Protocol Version 4 (TCP/IPv4) and click Properties.
Ethernet Properties dialog with Internet Protocol Version 4 (TCP/IPv4) selected and the Properties button highlighted
  1. Choose Use the following IP address and enter a static address on the rack's subnet:

Click OK.

IPv4 properties dialog with Use the following IP address selected and 192.168.0.100 / 255.255.255.0 / 192.168.0.1 entered
  1. Open a browser and navigate to 192.168.0.1. The Link+ Agilia configuration page loads. The browser will warn that the connection is not secure — this is expected for the device's local HTTP server.
Browser address bar showing 192.168.0.1 with the page titled Link+ Agilia and a not-secure warning
  1. On the authorized local setup connection, log in at the Identification — Password prompt (ID admin / password fresenius; if these are rejected, obtain the current credentials from Fresenius Kabi service).
Link+ Agilia web interface Identification - Password login form with ID and Password fields and a Login button
  1. Open the Configuration menu and select Data Export.
Link+ Agilia web interface Configuration menu expanded showing General Parameters, Network, Data Export, Time and Configuration Summary, with Data Export highlighted
  1. Under Serial export protocol for Agilia SP and VP, tick Enabled, then click Apply.
Data Export page with Serial export protocol for Agilia SP and VP set to Enabled, Serial export protocol Over TCP left disabled with port 52000, and the Apply button highlighted
  1. A dialog confirms the parameters were applied and warns that the Link+ will reboot when configuration is exited. Click OK.
Data Export confirmation dialog reading Applying parameters, please wait .. OK and warning that the settings will force the Link+ to reboot when configuration is exit
  1. Click Exit Configuration. The rack reboots on its own, which completes the setup.