Adding ADC to Device List #407

DGMC Clinical Investigations Facility (2021-05-17 21:50) · 549 view(s) · 2 comment(s)

Hello 

Many thanks for your great software. 
I am attempting to connect a Transonic Perivascular Flow Meter to Vital DB. I have purchased a DataQ DI-1100 ADC and made all of the connections to my laptop which uses Windows 7. With the Flow Meter in test mode, I receive a good signal using the WinDaQ software that comes with the ADC. However, when I attempt to add a device in VitalRecorder (v.1.8.15.15), nothing shows up in the COM port list. I have tried the instructions at the bottom of the add device page without success.  

Do you have any suggestions? 

Thank you!


vitaldb2017 (2021-05-23 22:37)
DATAQ instruments have two modes of operation.
When operating in the default libusb mode, it cannot be recognized by vital recorder.
You need to change to CDC mode according to the manufacturer's instructions below.
https://www.dataq.com/blog/data-acquisition/usb-daq-products-support-libusb-cdc

DGMC Clinical Investigations Facility (2021-05-27 18:34)
Thank you! That worked.