Elia Stanger (2022-07-21 12:25) · 456 view(s) · 1 comment(s)

Dear VitalDB-Team

I have some problems getting data out of the dräger primus. I have connected it and it is shown "on", but there is no data coming in. In addition to that, it often switches back to "off"  but then comes back to "on" seconds later. 

My question is if it is a problem in the machine's setup or if my configuration is false, however i can't really immagine that seeing there is not much to do wrong. 

Thanks for the help!


vitaldb2017 (2022-11-29 05:46)
Sorry for the late response.
Please check the device setting with the guide on page 38 of the document below.
https://www.draeger.com/Library/Content/IfU_Primus_SW_4.5n_EN_9053461.pdf
You should select the Medibus V4 and baud rate of 9.6K.