Carescape B850 #56

Jose De Jesus Lopez Rodriguez (2018-02-18 23:20) · 1927 view(s) · 1 comment(s)

Hi!!

Greetings and compliments on your software, is very interesting.

I would like some help, because i can not connect my GE carescape b850, that is mounted on a  GE carestation 650, I´ve tried everything, connecting the usb exit on the b850 to a null then to another serial - usb, I get the green light on the software but nothing more. Please help.

Thank you!


Chul-Woo Jung (2018-02-22 09:28)
You may refer the updated manual.
https://vitaldb.net/docs/document.php?document_id=1GFqZPtQhxEBd7hkLrTc7KIojxmklCtZi8zNsx6esk24§ion_id=h.ehr6ielsg94b

NOte: To communicate with a Carescape bx50 series monitor, you need two things: the ATEN UC-232A USB serial converter and a female-to-female null modem (cross gender)