Maquet / Getinge Servo-i / Servo-s / Servo-U Ventilators
Note: On Servo-i and Servo-s, connect to the BOTTOM RS-232 port. The TOP port is for service/debugging. The live port and adapter requirement for Servo-U have not been verified; confirm both with Getinge before connecting.
| Cable | Adapter | Port | Serial | VR Device Name |
|---|---|---|---|---|
| direct serial cable | Servo-i / Servo-s: Null Modem adapter (M/F); Servo-U: unverified | Servo-i / Servo-s: BOTTOM RS-232; Servo-U: unverified | Servo-i / Servo-s: 9600 baud, Even parity (7 or 8 data bits, 1 or 2 stop bits — auto-detected), XON/XOFF · Servo-U: 19200 | Servo-i |
Connection Steps
- Open the connector cover on the rear of the patient unit. On Servo-i and Servo-s, identify the BOTTOM of the two ports labeled RS 232; the upper port is for service/debugging.
- On Servo-i and Servo-s, attach a Null Modem adapter (M/F) to the bottom port.
- Connect a direct serial cable from the adapter to the PC via a USB-Serial converter.
Device Configuration
Nothing has to be enabled on the ventilator. The serial port is served by the built-in Computer Interface Emulator (CIE), which answers commands from Vital Recorder as soon as the line is up.
- Serial: 9600 baud, Even parity, XON/XOFF handshake. Data length (7 or 8 bits) and stop bits (1 or 2) are auto-detected by the CIE, so either works.
- The port is opto-isolated on both input and output, so a converter that relies on stealing power from the handshake lines may fail — use a properly powered USB-Serial converter.
Model differences: Servo-i and Servo-s run the CIE at 9600 baud; Servo-U runs at 19200 baud. Set the matching device entry in Vital Recorder.
Vital Recorder Setup
- In Vital Recorder, add the device as
Servo-i.