Dräger Infinity C500 / C700 (Infinity Acute Care System, IACS)
Note: This Infinity Acute Care System (IACS) connection records numeric data through the P2500 export port. ECG and arterial pressure waveforms require the Analog/Sync connection and an ADC.
| Connection | Cable or Interface | Adapter | Port | VR Device Name |
|---|---|---|---|---|
| Numeric data | Dräger export protocol cable MS22948 or custom RJ-10 to DB-9F cable | None for the cable described below | Export port on P2500 | Infinity |
| ECG/BP waveforms | Compatible Analog/Sync cable, connection cable for the ADC, and ADC | Depends on the ADC connection | Analog/Sync connection on M540 | Select the device type for your ADC |
Connection Requirements
Use the P2500 export port for this serial connection. No additional null modem adapter is required with the custom pin assignment below.
For analog waveforms, use a compatible Dräger Analog/Sync cable and ADC. If an Infinity M-Cable Microstream CO2 is also connected to the M540, use the compatible Y-cable arrangement for those accessories.
Cable Pinout
Numeric Data
| RJ-10 Pin (P2500 end) | DB-9F Pin (PC end) |
|---|---|
| 3 | 2 |
| 2 | 3 |
| 4 | 5 |
Connection Steps
Numeric Data
- Prepare a custom RJ-10 to DB-9F cable using the connections in Cable Pinout, or use the Dräger export protocol cable MS22948.
- Connect the RJ-10 end to the export port on the P2500.
- Connect the DB-9F end to the PC's serial port. If the PC has no serial port, use a USB-Serial converter.
ECG/BP Waveforms
Waveforms are only available as analog voltages, read through an ADC (SNU-ADC, DataQ DI-149/DI-155, …).
- Obtain a Dräger Analog/Sync cable.
- When the monitor is also using an Infinity M-Cable Microstream CO2, the Analog/Sync cable is joined through a Y cable and connected to the M540 monitor.
- Build a MDR14 ↔ RJ45 cable to bring the Analog/Sync cable's MDR connector into the ADC. The MDR-14 shell and the finished cable can both be ordered:
- Connect the ADC to the PC via USB.
- In Vital Recorder add the ADC (SNUADC or SNUADCM) and map the channels: ART = ch2 with gain ×100, ECG = ch3.
Device Configuration
No changes to the monitor settings are required for the numeric-data connection described here.
Vital Recorder Setup
Numeric Data
- Add the device as
Infinityand select the PC serial port used for the connection.
ECG/BP Waveforms
- Add the device type corresponding to your ADC.
- Configure the channels and signal scaling for the cable used. In the SNUADC example from this guide, ART uses channel 2 with gain ×100, and ECG uses channel 3. This mapping is specific to that cable and ADC setup.