Dräger Atlan

Dräger Atlan

Note: The Atlan speaks Dräger MEDIBUS.X and is added in Vital Recorder as MedibusX. The COM port must be set to MEDIBUS.X at 19200 before data arrives.

CableAdapterPortVR Device Name
Direct serial cableNull modem adapter (F/F)COM1 or COM2MedibusX

Connection Steps

  1. Locate the DB-9 serial port (COM 1 or COM 2) on the machine.
  2. Connect a direct serial cable to the COM1 or COM2 port using a Null Modem adapter (F/F).

Device Configuration

Open the machine's COM interface settings and configure the connected port as follows. Refer to the model's instructions for the menu path.

ParameterValue
ProtocolMEDIBUS.X
Baud Rate19200
Data Bits8
ParityEven
Stop Bits1

Vital Recorder Setup

Add the device as MedibusX and select the PC serial port used for the connection.