Dräger Perseus

Dräger Perseus

CableAdapterPortVR Device Name
Direct serial cableNull modem adapter (F/F)COM1 or COM2MedibusX

Connection Steps

  1. Attach a null modem adapter (F/F) to COM1 or COM2 on the machine's interface panel.
  2. Connect a direct serial cable between the null modem adapter and the PC's USB-Serial converter.

Use an available COM port for this connection. A Y-cable connection is not supported by this setup.

Device Configuration

  1. Open System setup → System.
  2. Enter the configuration password and select OK. The default password is 0000; use the site's current password if it has been changed.
  3. Select Interface.
  4. Set the connected port's Protocol to MEDIBUS.X and Baud rate to 19200.

The serial settings for this connection are:

ParameterValue
ProtocolMEDIBUS.X
Baud Rate19200
Data Bits8
ParityEven
Stop Bits1

Vital Recorder Setup

Add the device as MedibusX and select the PC serial port used for the connection.