Edwards Lifesciences Vigilance

Edwards Lifesciences Vigilance

CableAdapterPortVR Device Name
Direct serial cableNull modem adapter (M/F)COM 1 or COM 2Vigilance

Connection Steps

  1. Connect a direct serial cable to COM 1 using a null modem adapter (M/F). COM 2 can also be used; configure the port used for the connection.
Vigilance rear panel with the female DB-9 port labeled COM 1 circled, COM 2 beside it and the ANALOG IN 1/2 jacks to the right

Device Configuration

  1. Select Setup.
Vigilance home screen with the Trend, Patient Data, Setup and Alarms softkeys down the right edge, Setup circled
  1. Select System Config.
Vigilance Display Format screen with Home, Cursor and Change softkeys and the System Config softkey circled
  1. If the Patient Information screen appears, press Return to continue to configuration.
Vigilance Patient Information screen prompting for height, weight and BSA, with the Return softkey circled
  1. Select Digital Ports.
Vigilance System Configuration screen listing firmware revisions and the IFMout ID number, with the Digital Ports softkey circled
  1. Select the connected port and use Change to apply the settings below.
Vigilance Digital Ports screen with COM 1 set to Device IFMout, Baud Rate 9600, Parity None, Stop Bits 1, Data Bits 8 and Flow Control 2 seconds, while COM 2 Device is None

| Parameter | Value | |-----------|------- | | Device | IFMout | | Baud Rate | 9600 | | Parity | None | | Stop Bits | 1 | | Data Bits | 8 | | Flow Control | 2 seconds |

Vital Recorder Setup

Add the device as Vigilance and select the PC serial port used for the connection.