Edwards Lifesciences Vigileo

Edwards Lifesciences Vigileo

CableAdapterPortVR Device Name
Direct serial cableNull modem adapter (M/F)RS-232 serial portVigileo

Connection Steps

  1. Connect a direct serial cable to the rear serial port using a null modem adapter (M/F).
Vigileo rear panel with the recessed female DB-9 serial port circled, below the USB port

Device Configuration

  1. Tap the blank strip at the bottom left of the monitoring screen — there is no labeled button — to open the Status Menu.
Vigileo CO/SVV trend screen with the unlabelled blank strip at the bottom left circled
  1. In the Status Menu select Serial Port Setup.
Vigileo Status Menu listing New Patient, Patient Data, Display Setup, Serial Port Setup, Analog In Setup, Analog Out Setup, Default Settings, Engineering and Data Download, with Serial Port Setup highlighted
  1. Set Device to IFMout.
Vigileo Serial Port Setup screen with the Device picker open showing None, IFMout, Batch IFMout and Flexport, and IFMout highlighted
  1. Set Baud Rate to 9600, leave the other serial settings at their defaults, and press Return.
Vigileo Serial Port Setup screen with the Baud Rate picker open showing 1200, 2400, 9600, 19200, 38400 and 57600, and 9600 circled

Vital Recorder Setup

Add the device as Vigileo and select the PC serial port used for the connection.