LiDCO (LiDCOplus / LiDCOrapid)

LiDCO (LiDCOplus / LiDCOrapid)

Note: Serial output is off by default. The RS-232 communications module must be switched on inside the monitor's Engineering Screen before Vital Recorder will see anything, and the baud rate here is 57600 — not the 9600 used by the Edwards monitors.

CableAdapterPortSerialVR Device Name
direct serial cableNull Modem adapter (M/F) (assumes female COM1; unverified)COM1 — standard 9-way serial port on the monitor57600 baudLiDCO

Connection Steps

  1. Attach a Null Modem adapter (M/F) to COM1, the standard 9-way serial port on the monitor.
  2. Connect a direct serial cable from the adapter to the PC via a USB-Serial converter. The Null Modem supplies the TX/RX crossover; a direct serial cable alone will not work.

Device Configuration

  1. Enable the serial link and set its parameters in the monitor's Engineering Screen. The communications software module must also be switched on. The menu path is Settings → Communications → Serial.
  2. Set the following:

| Parameter | Value | |-----------|------- | | LiDCO Serial | Enabled | | Baud Rate | 57600 | | Data Bits | 8 | | Parity | None | | Stop Bits | 1 | | Average | Never | | Observation | Beat-to-beat |

  1. Average = Never and Observation = Beat-to-beat matter as much as the baud rate: with averaging enabled the monitor emits smoothed values at a slow cadence instead of per-beat data, and Vital Recorder's trend will look sparse or frozen.

Vital Recorder Setup

  • In the Vital Recorder device list this appears under Cardiac monitor as LiDCO: LiDCO.

Troubleshooting

  • Settings → Communications → Serial is not in the menu. Menu wording varies between LiDCOplus and LiDCOrapid and across software revisions. Look for the RS-232 port configuration inside the Engineering Screen.
  • Port opens but no data arrives. Check the adapter against the COM1 connector. If the port is male, use a Null Modem adapter (F/F). Confirm that the serial values match the table above.