Dräger EVITA (V300 / V500 / V600 / V800, Evita Infinity V500)

Dräger EVITA (V300 / V500 / V600 / V800, Evita Infinity V500)

Note: The EVITA family speaks Dräger MEDIBUS.X. The volume waveform is not transmitted, and CO2 requires a capnography module.

CableAdapterPortVR Device Name
direct serial cableNull modem; connector gender not confirmedRS-232 COM1 (or COM2)MedibusX

Connection Steps

  1. Connect the direct serial cable to the RS-232 COM port on the rear of the ventilator.

Device Configuration

Open the COM port / interface settings and configure the connected port as follows. The menu path varies by model and software version.

ParameterValue
ProtocolMEDIBUS.X
Baud Rate19200
Data Bits8
ParityEven
Stop Bits1

Vital Recorder Setup

  • Add the device as MedibusX and select the PC serial port used for the connection.

Troubleshooting

  • Vital Recorder restarts repeatedly when a MEDIBUS device is attached. A SIGSEGV loop present from 1.15.11 to 1.18.39 — fixed in 1.18.40; upgrade.