GE Corometrics 170 Series / 250cx

GE Corometrics 170 Series / 250cx

Note: The 170 Series and 250cx use different serial connectors and setup procedures. Use the section for your model.

ModelCableAdapterPortVR Device Name
170 SeriesCustom RJ-45 to DB-9F cable with RTS–CTS loopNoneRS-232 Port 1 or Port 2Coro
250cxCompatible RJ-11 serial cableConfirm for the cable usedRS-232C: J109, J110, or J111Coro

Connection Requirements

The custom cross cable provides the TX/RX crossover, so no separate Null Modem adapter is used.

Cable Pinout

Monitor RJ-45 port:

RJ-45 pinPort 1 signalPort 2 signalDirection
1+5 V (200 mA, fused)GND—
2RTSRTSOutput from monitor
3RXDRXDInput to monitor
4GNDGND—
5GNDGND—
6TXDTXDOutput from monitor
7CTSCTSInput to monitor
8+5 V (200 mA, fused)GND—

Pins 1 and 8 of Port 1 carry +5 V and are not connected to the PC side.

Cable wiring — three data conductors, crossed:

RJ-45 (monitor)SignalDB-9F (PC side)Signal
6TXD2RX
3RXD3TX
4 or 5GND5GND

Connect RJ-45 pin 7 (CTS) to RJ-45 pin 2 (RTS) at the monitor end.

Connection Steps

  1. Build the RJ-45 ↔ DB-9F cable as shown in Cable Pinout, including the CTS–RTS loop at the monitor end.
  2. Plug the RJ-45 end into RS-232 Port 1 (or Port 2) on the rear panel and the DB-9F end into the PC via a USB-Serial converter.

Corometrics 250cx

The 250cx has RJ-11 RS-232C ports (J109, J110 and J111), not the 170 Series' RJ-45 ports, so it needs a separate compatible RJ-11 serial cable. Confirm whether an adapter is required for the cable used. The 170 Series pinout above does not apply.

Device Configuration

The communication mode and baud rate for each port live in service setup mode, which can only be entered from a power-off state.

  1. Enter service setup mode:
    • Press and hold the Setup button (clock/calendar icon).
    • While still holding it, press and hold the blue Power button.
    • Release both buttons. Service mode is now active.
  2. Use the UA Reference button to toggle between the setup code (shown in the UA display) and its value (shown in the primary FHR display). The UA display is active when the ± qualifier is lit; the FHR display is active when the heartbeat indicator is lit.
  3. Use the Volume up / down buttons to change whichever display is active. On models 172, 173 and 174 use the leftmost set of volume controls.
  4. Set the following codes — only the port you actually cabled needs to be changed:

| Setup code (UA display) | Parameter | Value (FHR display) | |---|---|--- | | 30 | RS-232 Port 1 — communications mode | 5 (= 115 update) | | 31 | RS-232 Port 1 — baud rate | 9600 | | 40 | RS-232 Port 2 — communications mode | 5 (= 115 update) | | 41 | RS-232 Port 2 — baud rate | 9600 |

  • Communications mode values: 0 = HP, 1 = HP w/notes, 2 = ext. BP, 3 = factory test, 4 = ext. FSpO2, 5 = 115 update, 6 = 115 transmit/receive.
  • Baud rate values: 300, 600, 1200, 2400, 4800, 9600, 19200, 38400. The display abbreviates the larger values, so 9600 may appear as 96.
  • Factory defaults are not usable as-is: Port 1 ships as HP mode at 1200 baud, Port 2 as ext. BP at 600 baud.
  1. Exit with the Setup button. Exiting service setup mode puts the monitor into standby. If you exit with the Power button instead, none of your changes are saved.
  2. Serial: 9600 baud, port switched to communications mode 5 (115 update).

Vital Recorder Setup

  • Add the device in Vital Recorder as Coro.

Troubleshooting

  • The port is configured but nothing arrives. Verify the setup codes: 30/40 select the communications mode for ports 1/2 and 31/41 set their baud rate.

Notes

  • Recorded parameters: fetal HR1 / HR2 and uterine activity (UACT / TOCO). On current builds the maternal vital signs (SpO2, PR, NIBP) are also recorded into the same file as the fetal channels.
  • The 170 Series covers models 170, 171, 172, 173 and 174. The setup-code table above is common to all of them; the HR offset and ECG artifact elimination codes exist only on 172/173/174.