GE Dash 2000 / 3000 / 4000 / 5000
Note: This serial connection uses the AUX port and a custom RJ-45 to DB-9F cable. The separate Ethernet port is not used for this connection.
| Cable | Adapter | Port | VR Device Name |
|---|---|---|---|
| Custom RJ-45 to DB-9F cable | None | AUX (RJ-45) | Dashx000 |
Cable Pinout
| DB-9F Pin (PC end) | PC Signal | RJ-45 Pin (AUX end) | Monitor Signal |
|---|---|---|---|
| 2 | RxD | 6 | TxD |
| 3 | TxD | 3 | RxD |
| 5 | GND | 4 | GND |
Connection Steps
- Prepare a custom cable using the connections in Cable Pinout.
- Connect the RJ-45 end to the monitor's AUX port.
- Connect the DB-9F end to the PC's serial port. If the PC has no serial port, use a USB-Serial converter.
Device Configuration
No changes to the monitor settings are required.
Vital Recorder Setup
- Add the device as
Dashx000and select the PC serial port used for the connection.
Other Connections
- For ECG and arterial pressure through an ADC, see GE Defib Connectors. The Dash 4000 requires its own connector and ADC channel mapping.
- The GE Dash 2500 uses a different port and configuration procedure.