Note: Two ports, two jobs. Data is recorded from the 9-pin mini-DIN serial port at the bottom of the SpaceCom module, but the serial interface must first be enabled and its parameters set through the SpaceOnline web interface, reached over the LAN (RJ-45) port higher up on the same module.
| Cable | Adapter | Port | VR Device Name |
|---|---|---|---|
| Cable for 9-pin mini-DIN serial port | None | Serial (9-pin mini-DIN), bottom of the SpaceCom module | SpaceCom |
| Ethernet patch cable (configuration only) | None | LAN (RJ-45), SpaceOnline | — |
Connection Requirements
The SpaceCom has a 9-pin mini-DIN serial port and a separate LAN (RJ-45) port. Use the serial port for Vital Recorder data and the LAN port to configure the serial interface through SpaceOnline.
Connection Steps
- Identify the two ports on the SpaceCom module's connector column: the LAN port (SpaceOnline) near the top, below the two status LEDs, and the serial port (9-pin mini-DIN) at the bottom of the column.
- Connect a cable for the 9-pin mini-DIN serial port to the PC's serial interface or a USB-Serial converter.
Device Configuration
The serial interface is configured over LAN through SpaceOnline, the module's built-in web interface.
- Connect the SpaceCom LAN port to the PC with an Ethernet cable and give the PC a static address on the module's subnet:
- PC IP:
192.168.100.42/ Subnet:255.255.255.0/ Gateway:192.168.100.1 - The SpaceCom's own default Ethernet address is
192.168.100.41. - Open a browser to
192.168.100.41and log in. B. Braun ships three fixed accounts, each opening a different page — use the configuration one: status/status— status pageservice/service— service pageconfig/config— configuration page- Select Configuration in the left-hand menu. Set Language to English first if the interface came up in another language.
- Open BCC Protocol settings (its panel is titled Interface Settings) and set the serial parameters:
- Press Save. Power-cycle the SpaceStation so the interface comes up with the new settings.
- Interface: COM1
- Serial: 9600 baud, 8 data bits, No parity (
n), 1 stop bit
Vital Recorder Setup
- Add the device in Vital Recorder as
SpaceCom.