Fresenius Kabi Link+ Agilia
Note: Data is recorded through USB. Enable Data Export through the LAN connection before recording.
| Cable | Adapter | Port | VR Device Name |
|---|---|---|---|
| USB 2.0 A-male ↔ mini-B 5-pin | None | USB-B (mini-USB) | Link+ |
Connection Steps
- Connect the mini-B end to that port and the USB-A end directly to the PC. No USB-Serial converter is needed — the rack presents itself as a USB CDC-ACM serial device.
Device Configuration
Enabling Data Export is a one-time procedure performed over LAN from a PC. It must be done before the USB connection will produce any data.
- Connect the Link+ LAN (RJ-45) port — on the left of the same connector panel — to the PC's Ethernet port with the LAN cable supplied with the Link+.
- On the PC, open Control Panel → Network and Internet → Network Connections, right-click the Ethernet adapter and choose Properties.
- In the adapter properties list, select Internet Protocol Version 4 (TCP/IPv4) and click Properties.
- Choose Use the following IP address and enter a static address on the rack's subnet:
- IP address:
192.168.0.100 - Subnet mask:
255.255.255.0 - Default gateway:
192.168.0.1
Click OK.
- Open a browser and navigate to
192.168.0.1. The Link+ Agilia configuration page loads. The browser will warn that the connection is not secure — this is expected for the device's local HTTP server.
- On the authorized local setup connection, log in at the Identification — Password prompt (ID
admin/ passwordfresenius; if these are rejected, obtain the current credentials from Fresenius Kabi service).
- Open the Configuration menu and select Data Export.
- Under Serial export protocol for Agilia SP and VP, tick Enabled, then click Apply.
- A dialog confirms the parameters were applied and warns that the Link+ will reboot when configuration is exited. Click OK.
- Click Exit Configuration. The rack reboots on its own, which completes the setup.
Vital Recorder Setup
- Add a device with Device Type
Link+and leave Port empty. Vital Recorder finds the rack by its USB ID and follows it when it re-connects under a new port number. Leave Y Cable unchecked.
Troubleshooting
- Crashes or a failed reconnection on an older build. Several Link+-specific crashes and a reconnection failure were fixed across 1.18.50–1.19.5; upgrade before troubleshooting hardware.